Article 4.11.1F : Programming range and Full scale display of ADC & VDC meter.
Please find the below image of the backside of ADC & VDC. the input of ADC and VDC is shown in SR No sticker it should be selected by the customer as per requirement. Vin or I/P will be shown in Sr no sticker.
Standard input and programming range is given below also you can see meter backside Image. Standard input was given on Sr No description like Vin or I/P.
Standard
Input |
Programming
Range |
10V |
1-10 V |
30V |
10-40 V |
100V |
40-150 V |
400V |
150-500 V |
20mA |
0-20 Ma |
75mV |
50-100 mV |
Full-scale display: It is used for other than input we can program display reading with this formula. Input is the same we cannot change input it's set by the factory during manufacture. Please find the FS Display formula below.
FS = Standard Input x Display FS
Actual Input